FDA Adverse Event Malfunction Summary report: N

LOANER CORE SAG SAW

MDR report key: 3072845 · Received April 23, 2013

Report

Report Number
0001811755-2013-00870
Event Type
Malfunction
Date Received
April 23, 2013
Date of Event
March 27, 2013
Report Date
March 27, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON EVALUATION, THE MOTOR AND PRESS PLUG WERE REPLACED DUE TO THE ERROR MESSAGE. PREVENTATIVE MAINTENANCE WAS PERFORMED AND DUE TO THE DEVICE LOANER STATUS, THIS DOES NOT REQUIRE RETURN.

Description of Event or Problem · 1

THE CORE SAG SAW WAS RETURNED AFTER THE ACCOUNT RECEIVED THEIR OWN DEVICE BACK FROM REPAIR. UPON EVALUATION AT THE MANUFACTURER, IT DISPLAYED A BIAS CURRENT MESSAGE WHEN CONNECTED TO THE CONSOLE SIGNALING A CONDITION OCCURED FROM WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
175667 LOANER CORE SAG SAW DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1