SYNCHROMED II
Report
- Report Number
- 3004209178-2013-06743
- Event Type
- Injury
- Date Received
- April 23, 2013
- Report Date
- March 27, 2013
- Manufacturer
- MDT PUERTO RICO OPERATIONS CO
- Product Code
- LKK
- PMA / PMN Number
- P860004
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT ID: 8711, SERIAL# (B)(4), IMPLANTED: (B)(6) 2011, PRODUCT TYPE: CATHETER. (B)(4).
IT WAS REPORTED THAT THE PATIENT HAD INCREASED TIGHTNESS THAT STARTED APPROXIMATELY IN NOVEMBER AND THE SYMPTOMS WOULD COME AND GO. THE SITE OVER THE PUMP WAS SWOLLEN FOR APPROXIMATELY THE PAST WEEK. A CATHETER ACCESS PORT (CAP) ASPIRATION AND DYE STUDY WERE PERFORMED AND THE RESULTS WERE REPORTED AS "EVERYTHING SEEMED OKAY." IT WAS NOTED THAT THE LAST HEALTH CARE PROVIDER (HCP) THAT EXAMINED THE PATIENT WAS NERVOUS ABOUT FILLING THE PUMP DUE TO THE SWELLING, AND THOUGHT THERE WAS SOME KIND OF DISCONNECT. AN X-RAY WAS PERFORMED AND IT WAS NOTED IT LOOKED "OKAY". THERE WAS TRANSLUCENCY NOTED WHERE THE CATHETERS CONNECTED, AND AFTER THE TRANSLUCENCY THE CATHETER LOOKED A LITTLE "BLURRED AND DIFFERENT"; HOWEVER, THE PHYSICIAN BELIEVED IT MAY HAVE BEEN DUE TO THE WAY IT WAS JOINED. THE PHYSICIAN STATED IT LOOKED LIKE A SEROMA. AN ABDOMINAL BINDER WAS RECOMMENDED TO HELP WITH THE SWELLING. THE PUMP WAS DELIVERING GABLOFEN. IT WAS LATER REPORTED THAT THE EVENT WAS DUE TO DRUG-EFFECTS. IT WAS NOTED THE PATIENT SEEMED TO BE IN WITHDRAWAL, AND HAD EXPERIENCED INCREASED TIGHTNESS IN THE LEGS. A CAP ASPIRATION UNDER FLUOROSCOPY WITH DYE WAS PERFORMED ON (B)(6) 2013, WHICH SHOWED NORMAL FLOW TO THE INTRATHECAL SPACE. AN X-RAY WAS INDICATED TO HAVE BEEN DONE ON (B)(6) 2013. IT WAS ALSO NOTED THAT A BOLUS WAS GIVEN AND PATIENT RECOVERED WITHOUT SEQUELA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 174899 | SYNCHROMED II | PUMP, INFUSION, IMPLANTED, PROGRAMMABLE | LKK | MDT PUERTO RICO OPERATIONS CO | 8637-20 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00014 YR | Other |