FDA Adverse Event Injury Summary report: N

ALTRX NEUT 36IDX58OD

MDR report key: 3072821 · Received April 23, 2013

Report

Report Number
1818910-2013-15753
Event Type
Injury
Date Received
April 23, 2013
Date of Event
April 9, 2013
Report Date
April 9, 2013
Manufacturer
1818910 DEPUY ORTHOPAEDICS, INC.
Product Code
LPH
PMA / PMN Number
PK072963
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATIONS SINCE THEIR RELEASE FOR DISTRIBUTION. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A; REV. D. OPERATIVE NOTES AND PATHOLOGY REPORTS WERE OBTAINED BUT DO NOT CONFIRM THE INFECTION OR IDENTIFY A ROOT CAUSE. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS DEEP INFECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
173639 ALTRX NEUT 36IDX58OD ACETABULAR LINER LPH 1818910 DEPUY ORTHOPAEDICS, INC. 338073

Patients

Seq Age Sex Outcome Treatment
1 53 YR Required Intervention