VIPER2 X-TAB CANNULATED POLYAXIAL SCREWDRIVER
Report
- Report Number
- 1526439-2013-15723
- Event Type
- Malfunction
- Date Received
- April 23, 2013
- Date of Event
- March 27, 2013
- Report Date
- April 22, 2013
- Manufacturer
- DEPUY SYNTHES SPINE
- Product Code
- HWR
- PMA / PMN Number
- PEXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
A FOLLOW UP REPORT WILL BE FILED UPON COMPLETION OF THE INVESTIGATION.
VISUAL EXAMINATION OF THE RETURNED DRIVER FOUND THAT THE MOST DISTAL PORTION OF THE TIP HAD BROKEN OFF FROM THE INSTRUMENT. REVIEW OF DEVICE HISTORY RECORDS CONFIRMED THE INSTRUMENT WAS MANUFACTURED TO SPECIFICATION REQUIREMENTS. NO DEFINITIVE CONCLUSIONS CAN BE MADE AS TO THE CAUSE OF TIP BREAKAGE. HOWEVER, A CAPA WAS IMPLEMENTED WHICH ADDRESSED TIP BREAKAGE THROUGH DESIGN MODIFICATION AND MATERIAL CHANGE. TESTING ON PRODUCTION LEVEL INSTRUMENTS HAD DETERMINED THAT TIP BREAKAGE WAS THE RESULT OF A BRITTLE FRACTURE OF THE MATERIAL. THIS PRODUCTION LOT WAS MANUFACTURED PRIOR TO THOSE CHANGES. AT THIS TIME, THE COMPLAINT IS CONSIDERED TO BE CLOSED.
INTERNATIONAL AFFILIATE REPORTS THAT INSPECTION OF A RETURNED LOANER KIT FOUND THE TIP OF THE POLYAXIAL DRIVER HAD BROKEN OFF FROM THE INSTRUMENT. IT IS NOT KNOWN WHEN/WHERE TIP BREAKAGE OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 174772 | VIPER2 X-TAB CANNULATED POLYAXIAL SCREWDRIVER | DRIVER, PROSTHESIS | HWR | DEPUY SYNTHES SPINE | MI19980 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |