FDA Adverse Event Malfunction Summary report: N

ISOLA SET SCW HEXDRIVER

MDR report key: 3072492 · Received April 23, 2013

Report

Report Number
1526439-2013-15719
Event Type
Malfunction
Date Received
April 23, 2013
Date of Event
March 27, 2013
Report Date
April 24, 2013
Manufacturer
DEPUY SYNTHES SPINE
Product Code
LXH
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FOLLOW UP REPORT WILL BE FILED UPON RECEIPT OF THE DEVICE AND COMPLETION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 1

A VISUAL INSPECTION NOTED THAT HEX TIP OF THE DRIVER HAD SNAPPED OFF FROM THE INSTRUMENT. THE BROKEN HEX TIP SECTION WAS NOT RETURNED FOR FURTHER EVALUATION. REVIEW OF THE DEVICE HISTORY RECORD CANNOT BE PERFORMED AS THE LOT CODE IS UNKNOWN. NO DEFINITIVE CONCLUSIONS CAN BE MADE. HOWEVER, THE NOTED DAMAGE SUGGESTS THAT THE DEVICE HEX TIP MOST LIKELY BROKE OFF DUE TO AN UNANTICIPATED TORSIONAL STRESS. IN THE ABSENCE OF A PRODUCT ISSUE OR OBSERVED TREND, THIS COMPLAINT WILL BE CLOSED WITH NO FURTHER ACTION REQUIRED.

Description of Event or Problem · 1

INTERNATIONAL AFFILIATE REPORTS THAT INSPECTION OF A RETURNED LOANER KIT FOUND THE TIP OF THE HEX DRIVER HAD BROKEN OFF FROM THE INSTRUMENT. IT IS NOT KNOWN WHEN/WHERE TIP BREAKAGE OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
173230 ISOLA SET SCW HEXDRIVER ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH DEPUY SYNTHES SPINE UNK

Patients

Seq Age Sex Outcome Treatment
1