FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3100 SYSTEM
MDR report key: 3072318
·
Received April 5, 2013
Report
- Report Number
- 3008642652-2013-00908
- Event Type
- Malfunction
- Date Received
- April 5, 2013
- Date of Event
- March 7, 2013
- Report Date
- April 2, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WRENCH CODE 6) HAS BEEN CONFIRMED. UPON INVESTIGATION, THE FRONT RESPONSE BUTTON WAS STUCK, WHICH CAUSED THE WRENCH CODE REPORTED BY THE PT. THE ROOT CAUSE OF THE STUCK RESPONSE BUTTON WAS UNABLE TO BE POSITIVELY DETERMINED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE RESPONSE BUTTON. THE PT RECEIVED A REPLACEMENT MONITOR.
Description of Event or Problem · 1
A (B)(6) FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HER MONITOR WAS DISPLAYING A WRENCH CODE 6 (RESPONSE BUTTON STUCK). THE PT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 141864 | LIFEVEST WCD 3100 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |