ANIMAS INSULIN INFUSION PUMP
MDR report key: 3072264
·
Received April 23, 2013
Report
- Report Number
- 2531779-2013-04860
- Event Type
- Malfunction
- Date Received
- April 23, 2013
- Report Date
- March 27, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
(B)(6). SUBMITTED: 04/23/2013 DEVICE EVALUATION: THE INSULIN PUMP HAS BEEN RETURNED AND ADDITIONAL INVESTIGATION WAS PERFORMED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: ON EXAMINATION, THERE WAS NO VISIBLE DAMAGE TO THE KEYPAD. ON INVESTIGATION, THE CONTRAST BUTTON WAS UNRESPONSIVE. THE REMAINDER OF THE KEYPAD BUTTONS HAD NORMAL RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION PRESENT UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS.
Description of Event or Problem · 1
THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013: THERE WAS CONTAMINATION UNDER THE CONTACTS OF ALL THE BUTTONS.