FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 3072257 · Received April 23, 2013

Report

Report Number
2531779-2013-04857
Event Type
Malfunction
Date Received
April 23, 2013
Report Date
March 26, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION (B)(4) 2013 DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A RESERVED SAMPLE FROM THE SAME LOT NUMBER WAS TESTED. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED WITH NO DAMAGE OR DEFECTS NOTED. A LEAK, FORCE AND FILL TEST WAS PERFORMED WITH NO FAILURES. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.

Additional Manufacturer Narrative · 1

FOLLOW-UP #2 DATE OF SUBMISSION 08/06/2013, DEVICE EVALUATION: THE CARTRIDGE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED WITH NO DAMAGE OR DEFECTS NOTED. A LEAK, FORCE AND FILL TEST WAS PERFORMED WITH NO FAILURES.

Additional Manufacturer Narrative · 1

(B)(6). THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE SUBJECT CARTRIDGE(S) HAS NOT BEEN RETURNED TO ANIMAS. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A SITE/SET/CART (CARTRIDGE ISSUE) ISSUE; MOISTURE IN THE CARTRIDGE COMPARTMENT FOR THE PAST 15 CARTRIDGE CHANGES. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT AS THE REPORTED BLOOD GLUCOSE EXCURSIONS DID NOT MEET ANIMAS' CRITERIA FOR AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
174947 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION B201805

Patients

Seq Age Sex Outcome Treatment
1 40 YR