FDA Adverse Event Malfunction Summary report: N

THERMOPHORE AUTOMATIC MOIST HEAT PACK

MDR report key: 3072224 · Received September 10, 2012

Report

Report Number
1811605-2012-00143
Event Type
Malfunction
Date Received
September 10, 2012
Date of Event
September 1, 2012
Report Date
September 4, 2012
Manufacturer
BATTLE CREEK EQUIPMENT COMPANY
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED THAT THE SWITCH WOULD NOT TURN OFF. UPON, EXAMINATION, WE DISCOVERED THAT A RESISTOR SHORTED ON THE CIRCUIT BOARD. THE FLAME-RETARDANT SWITCH HOUSING CONTAINED THE EVENT PROPERLY. AT OUR REQUEST, THE MANUFACTURER OF THE SWITCH HAS ENACTED SEVERAL CAPA PROJECTS TO IMPROVE THE QUALITY OF THE SWITCH. THE OCCURRENCE OF BOARD COMPONENT FAILURE HAS DROPPED SIGNIFICANTLY IN THE PAST 12 MONTHS.

Description of Event or Problem · 1

IT WAS REPORTED THE SWITCH WOULD NOT TURN OFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERMOPHORE AUTOMATIC MOIST HEAT PACK ELECTRIC MOIST HEAT PACK IRT BATTLE CREEK EQUIPMENT COMPANY 055

Patients

Seq Age Sex Outcome Treatment
1