FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE AUTOMATIC MOIST HEAT PACK
MDR report key: 3072224
·
Received September 10, 2012
Report
- Report Number
- 1811605-2012-00143
- Event Type
- Malfunction
- Date Received
- September 10, 2012
- Date of Event
- September 1, 2012
- Report Date
- September 4, 2012
- Manufacturer
- BATTLE CREEK EQUIPMENT COMPANY
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
IT WAS REPORTED THAT THE SWITCH WOULD NOT TURN OFF. UPON, EXAMINATION, WE DISCOVERED THAT A RESISTOR SHORTED ON THE CIRCUIT BOARD. THE FLAME-RETARDANT SWITCH HOUSING CONTAINED THE EVENT PROPERLY. AT OUR REQUEST, THE MANUFACTURER OF THE SWITCH HAS ENACTED SEVERAL CAPA PROJECTS TO IMPROVE THE QUALITY OF THE SWITCH. THE OCCURRENCE OF BOARD COMPONENT FAILURE HAS DROPPED SIGNIFICANTLY IN THE PAST 12 MONTHS.
Description of Event or Problem · 1
IT WAS REPORTED THE SWITCH WOULD NOT TURN OFF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMOPHORE AUTOMATIC MOIST HEAT PACK | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT COMPANY | 055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |