OT VERIO IQ METER
Report
- Report Number
- 3008382007-2013-08709
- Event Type
- Malfunction
- Date Received
- April 23, 2013
- Report Date
- March 27, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Removal / Correction Number
- 2013-IE-017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
FOLLOW-UP # 1 ¿ (6/16/2013) THE PATIENT¿S METER AND TEST STRIPS HAVE BEEN RETURNED ON (B)(4) 2013 AND (B)(4) 2013, RESPECTIVELY, AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 AND (B)(4) 2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS: THE METER AND TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER WAS FOUND TO BE UNABLE TO REPRODUCE AN ERROR 2 MESSAGE. THE TEST STRIPS WERE EVALUATED AND A SECONDARY ISSUE WAS NOTED WHERE ERROR 4 WAS OBSERVED DURING CONTROL SOLUTION TESTS. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING ERROR UNKNOWN. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 173341 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3376973 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |