ADULT DUAL-HEATED WITH EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2013-00320
- Event Type
- Malfunction
- Date Received
- April 22, 2013
- Date of Event
- March 26, 2013
- Report Date
- March 26, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- REFER TO H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: THE COMPLAINT RT340 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH) AND VISUALLY INSPECTED. RESULTS: A HOLE WAS FOUND IN THE EVAQUA (EXPIRATORY) LIMB OF THE COMPLAINT BREATHING CIRCUIT, APPROXIMATELY 7 CM AWAY FROM THE PATIENT END CONNECTOR. CONCLUSION: BASED ON INSPECTION OF THE DAMAGE, IT IS LIKELY THAT THE EVAQUA EXPIRATORY LIMBS WERE PUNCTURED OR SCRATCHED BY A BLUNT OBJECT. ALL RT340 BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS PRIOR TO DISTRIBUTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THAT THE BREATHING CIRCUIT WAS DAMAGED POST-PRODUCTION, POSSIBLY DURING STORAGE OR SETUP. (B)(4). THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 BREATHING CIRCUIT STATE: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; SET APPROPRIATE VENTILATOR ALARMS; FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE.
A HOSPITAL IN (B)(6) REPORTED THAT A HOLE WAS FOUND ON THE EXPIRATORY LIMB OF AN RT340 ADULT BREATHING CIRCUIT. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 171825 | ADULT DUAL-HEATED WITH EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT340 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|