FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED CHAMBER

MDR report key: 3072071 · Received April 22, 2013

Report

Report Number
9611451-2013-00319
Event Type
Malfunction
Date Received
April 22, 2013
Date of Event
March 26, 2013
Report Date
March 26, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290 CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) AND VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THERE WAS A BREAK IN THE FEEDSET TUBE WHERE IT IS INSERTED INTO THE CHAMBER. THE SURFACE OF THE BREAK IS ROUGH AND NOT SMOOTHLY CUT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS TYPE FOR LOT NUMBER 120325. CONCLUSION: THE DAMAGE APPEARED TO BE DUE TO THE FEEDSET TUBE BEING PULLED, AWAY FROM THE CHAMBER, POSSIBLY FROM THE FEEDSET BEING CAUGHT OR UNDER TENSION. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THE USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT A WATER LEAK OCCURRED AT THE CONNECTION BETWEEN THE WATER FEEDSET TUBE AND AN MR290 AUTOFEED HUMIDIFICATION CHAMBER DOME. THIS WAS FOUND PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172567 VENTED AUTOFEED CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 1203250304

Patients

Seq Age Sex Outcome Treatment
1