FDA Adverse Event Malfunction Summary report: N

5MM, 33CM PEEK MONOPOLAR HANDLE 33CM

MDR report key: 3071927 · Received April 22, 2013

Report

Report Number
0002936485-2013-00169
Event Type
Malfunction
Date Received
April 22, 2013
Date of Event
April 1, 2013
Report Date
April 1, 2013
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
KOG
PMA / PMN Number
K973259
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED FOR INVESTIGATION. THE REPORTED FAILURE MODE WAS CONFIRMED. THE INSULATION IS CRACKED AND MELTED NEAR THE DISTAL TIP. THE INSULATION WAS TESTED FOR CRACKS/DAMAGES AND THE UNIT FAILED THE INSULATION SCAN TEST. THE PROBABLE ROOT CAUSE IS USER MISUSE. IN SUM, THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION OF THE DEVICE WAS COMPROMISED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION OF THE DEVICE WAS COMPROMISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172911 5MM, 33CM PEEK MONOPOLAR HANDLE 33CM ENDOSCOPE AND/OR ACCESSORIES KOG STRYKER ENDOSCOPY-SAN JOSE 1040655D

Patients

Seq Age Sex Outcome Treatment
1