FDA Adverse Event
Malfunction
Summary report: N
5MM, 33CM PEEK MONOPOLAR HANDLE 33CM
MDR report key: 3071914
·
Received April 22, 2013
Report
- Report Number
- 0002936485-2013-00167
- Event Type
- Malfunction
- Date Received
- April 22, 2013
- Date of Event
- April 1, 2013
- Report Date
- April 1, 2013
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- KOG
- PMA / PMN Number
- K973259
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PRODUCT WAS RETURNED FOR INVESTIGATION. THE REPORTED FAILURE MODE WAS CONFIRMED. THE INSULATION IS MELTED NEAR THE DISTAL TIP AND DENTS WERE ALSO SEEN. THE INSULATION WAS TESTED FOR CRACKS/DAMAGES AND THE UNIT FAILED THE INSULATION SCAN TEST. THE PROBABLE ROOT CAUSE IS USER MISUSE. IN SUM, THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.
Additional Manufacturer Narrative · 1
ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULATION OF THE DEVICE WAS COMPROMISED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULATION OF THE DEVICE WAS COMPROMISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 172839 | 5MM, 33CM PEEK MONOPOLAR HANDLE 33CM | ENDOSCOPE AND/OR ACCESSORIES | KOG | STRYKER ENDOSCOPY-SAN JOSE | ¿1040667D¿ |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |