FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 3071858 · Received April 22, 2013

Report

Report Number
1644487-2013-01102
Event Type
Malfunction
Date Received
April 22, 2013
Date of Event
November 19, 2008
Report Date
March 24, 2013
Manufacturer
CYBERONICS
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF PROGRAMMING HISTORY.

Description of Event or Problem · 1

THROUGH REVIEW OF PROGRAMMING HISTORY A DEFAULTED GENERATOR DIAGNOSTICS ON (B)(6) 2008 CAUSED THE GENERATOR TO CHANGE ITS SETTINGS TO UNDESIRABLE VALUES. ON VISIT (B)(6) 2008, THE PATIENT WAS PROGRAMMED AT SETTINGS OUTPUT CURRENT= 1.75 MA/ FREQUENCY= 20 HZ/ PULSE WIDTH= 250 USEC/ON TIME= 60 SEC/OFF TIME= 1.1 MIN / MAGNET OUTPUT CURRENT= 2 MA/ ON TIME= 60 SEC/ PULSE WIDTH= 250 USEC. A SYSTEM DIAGNOSTICS WAS THEN PERFORMED ON THE PATIENT'S VNS GENERATOR WHICH WAS FAULTED AND CHANGED THE PATIENT'S SETTINGS. NO INTERROGATION WAS PERFORMED BEFORE THE PATIENT LEFT THE OFFICE. DURING THE NEXT VISIT ON (B)(6) 2008, THE PATIENT'S VNS GENERATOR WAS INTERROGATED AND WAS FOUND AT OUTPUT CURRENT= 1 MA/ FREQUENCY= 20 HZ/ PULSE WIDTH= 500 USEC/ON TIME= 30 SEC/OFF TIME= 60 MIN / MAGNET OUTPUT CURRENT= 1 MA/ ON TIME= 500 SEC/ PULSE WIDTH= 30 USEC. THE PATIENT'S SETTINGS WERE THEN CHANGED TO OUTPUT CURRENT= 2 MA/ FREQUENCY= 20 HZ/ PULSE WIDTH= 250 USEC/ON TIME= 60 SEC/OFF TIME= 1.1 MIN / MAGNET OUTPUT CURRENT= 2.25 MA/ ON TIME= 60 SEC/ PULSE WIDTH= 250 USEC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172728 PROGRAMMING SOFTWARE PROGRAMMING SOFTWARE LYJ CYBERONICS MODEL 250 521380

Patients

Seq Age Sex Outcome Treatment
1 27 YR