FDA Adverse Event Injury Summary report: N

TALENT AAA

MDR report key: 3071838 · Received April 22, 2013

Report

Report Number
2953200-2013-00752
Event Type
Injury
Date Received
April 22, 2013
Date of Event
February 3, 2012
Report Date
June 27, 2013
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P070027
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). METHODS: (FILMS); RESULTS: INHERENT RISK OF PROCEDURE (ENDOLEAK). CONCLUSION: (INSUFFICIENT INFORMATION, CAUSE OF EVENTS IS UNKNOWN); KNOWN INHERENT RISK OF PROCEDURE (ENDOLEAK).

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. THE ANEURYSM DIAMETER WAS 5.4 CM. ADDITIONAL ANEURYSM AND VESSEL MORPHOLOGY INFORMATION WAS NOT REPORTED. AT THE INDEX PROCEDURE THE PATIENT RECEIVED A TALENT (B)(4) BIFURCATED STENT GRAFT AND AN ANEURX (B)(4) LIMB. APPROXIMATELY THREE YEARS AFTER THE INDEX PROCEDURE, THE PATIENT PRESENTED WITH A PROXIMAL TYPE I ENDOLEAK. ENDURANT (B)(4) CUFFS WERE IMPLANTED, AND THE ENDOLEAK RESOLVED. APPROXIMATELY FOUR YEARS AFTER THE INDEX PROCEDURE, THE PATIENT PRESENTED WITH A TYPE III FABRIC ENDOLEAK. AN ENDURANT (B)(4) LIMB WAS IMPLANTED, AND THE ENDOLEAK RESOLVED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE. REVIEW OF A SINGLE STILL ANGIO IMAGE SHOWED CONTRAST OUTSIDE OF THE STENTLESS AREA OF THE IPSILATERAL LIMB; POSSIBLY DUE TO A TYPE III FABRIC ENDOLEAK. AN IMAGE AFTER RELINING WITH A LIMB SHOWED RESOLUTION OF THE ENDOLEAK. THE CAUSE OF THE ENDOLEAK COULD NOT BE DETERMINED FROM THIS IMAGE.

Description of Event or Problem · 1

A REVIEW OF CTA'S 4 YEARS POST-IMPLANT SHOW THAT THE ENDURANT CUFF(S) WERE POSITIONED JUST BELOW THE RENAL ARTERIES. THE TALENT BIFURCATE WAS 2-3CM BELOW THE RENALS; THE ORIGINAL PLACEMENT OF THE TALENT IS UNKNOWN. THE AAA WAS 8CM MAX DIAMETER. THE IPSILATERAL LIMB WAS PLACED WITHIN THE LEFT ILIAC. THE AORTIC BODY WAS ANGULATED 90DEG TO THE LIMBS (A-P). BELOW THE ANGULATION, A POSSIBLE TYPE III FABRIC ENDOLEAK IS SEEN NEAR THE STENTLESS AREA OF THE IPSILATERAL LIMB IN THE DISTAL AAA SAC. THERE IS A LARGE AREA OF CALCIFICATION NEAR THE ENDOLEAK IN THE DISTAL AAA. THE CAUSE COULD NOT BE DETERMINED. IT IS POSSIBLE THAT CALCIFICATION MAY HAVE ABRADED THE FABRIC AND CAUSED THE TYPE III ENDOLEAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
171846 TALENT AAA SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC CARDIOVASCULAR V00183523

Patients

Seq Age Sex Outcome Treatment
1 00090 YR Required Intervention