FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY

MDR report key: 3071807 · Received April 22, 2013

Report

Report Number
2649622-2013-05287
Event Type
Injury
Date Received
April 22, 2013
Date of Event
February 14, 2013
Report Date
February 14, 2013
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4) IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) (B)(6) 2013; (B)(4) IMPLANTABLE PACING LEAD (B)(6) 2013; (B)(4) IMPLANTABLE TACHY LEAD (B)(6) 2013. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT APPROXIMATELY SEVEN DAYS AFTER IMPLANT, THERE WAS NO SENSING AND LOW LEAD IMPEDANCE THROUGH THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) ON THE RIGHT ATRIAL (RA) LEAD, THE RIGHT VENTRICULAR (RV) LEAD, AND THE LEFT VENTRICULAR (LV) LEAD. ALL MEASUREMENTS APPEARED TO BE NORMAL FOR ALL THREE LEADS THROUGH THE ANALYZER. ALSO, THE ICD BATTERY VOLTAGE DIPPED TO BELOW END OF LIFE (EOL) LEVELS FOR THREE CONSECUTIVE DAILY MEASUREMENTS WITHOUT TRIGGERING EOL OPERATION. THE ICD WAS EXPLANTED AND REPLACED. ALL THREE LEADS REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
171810 ATTAIN ABILITY DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 419688

Patients

Seq Age Sex Outcome Treatment
1 00051 YR Hospitalization| R