FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3071584 · Received April 22, 2013

Report

Report Number
2531779-2013-04818
Event Type
Malfunction
Date Received
April 22, 2013
Report Date
March 27, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 07/01/2013- DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX HISTORY SHOWS NO EVIDENCE OF A ¿LOAD STEP MALFUNCTION.¿ ONE ¿NO CARTRIDGE DETECTED¿ WARNING WAS OBSERVED ON THE REPORTED DATE AS A RESULT OF A LOAD ATTEMPT AFTER AN ¿EMPTY CARTRIDGE¿ WARNING WITHOUT REWINDING. DURING TESTING THE PUMP POWERED ON AND WAS PRIMED SUCCESSFULLY; NO ¿LOAD STEP MALFUNCTION¿ WAS DUPLICATED. THE FORCE SENSOR CALIBRATION WAS FOUND TO BE WITHIN SPECIFICATIONS. THE PUMP WAS OPENED AND NO DEFECT WAS FOUND.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOAD STEP MALFUNCTION) ISSUE. THE PUMP REPORTEDLY DID NOT RECOGNIZE A FILLED CARTRIDGE AND DISPENSED INSULIN DURING THE LOAD CARTRIDGE STEP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172954 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 12 YR