FDA Adverse Event Malfunction Summary report: N

SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION

MDR report key: 3071569 · Received April 22, 2013

Report

Report Number
2015691-2013-19880
Event Type
Malfunction
Date Received
April 22, 2013
Date of Event
March 26, 2013
Report Date
March 26, 2013
Manufacturer
EDWARDS LIFESCIENCES, PR
Product Code
LDF
PMA / PMN Number
K813521
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ONE BIPOLAR PACING CATHETER WITH ATTACHED MONOJECT 1.3CC LIMITED VOLUME SYRINGE WAS RETURNED FOR EVALUATION. NO INTRODUCER WAS RETURNED WITH THE CATHETER. CONTINUITY TESTING FOUND THAT A SHORT CONDITION OCCURRED IN THE Y-ADAPTOR BETWEEN THE PROXIMAL AND DISTAL LEADWIRES. NO OPEN OR SHORT CONDITIONS WERE OBSERVED IN THE LEADWIRES BETWEEN THE DISTAL PORTION OF THE Y-ADAPTOR AND THE ELECTRODES. THE BALLOON INFLATED CLEAR AND CONCENTRIC AND REMAINED INFLATED FOR 5 MINUTES WITHOUT LEAKAGE. NO VISIBLE DAMAGE TO THE CATHETER BODY, BALLOON, WINDINGS, OR RETURNED SYRINGE WAS OBSERVED. CONTINUITY TESTING WAS DONE BY CONNECTING ONE TEST LEAD OF A MULTIMETER TO AN ELECTRODE CONNECTOR PIN AND THE OTHER LEAD TO THE RELATED ELECTRODE. RESISTANCE WAS MEASURED WITH DIGITAL MULTIMETER. BALLOON INFLATION TESTING WAS PERFORMED USING RETURNED SYRINGE WITH 1.3CC AIR BY HOLDING THE BALLOON UNDER WATER. VISUAL EXAMINATIONS WERE PERFORMED UNDER MICROSCOPE AT 10X MAGNIFICATION. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED AND DOCUMENTED THAT THE DEVICE MET ALL SPECIFICATIONS UPON DISTRIBUTION. THE CUSTOMER REPORT OF PACING DIFFICULTY WAS CONFIRMED DURING THE EVALUATION. AN INVESTIGATION HAS BEEN INITIATED TO DETERMINE IF THE ROOT CAUSE IS RELATED TO THE MANUFACTURING PROCESS AND IMPLEMENT ANY NECESSARY CORRECTIVE ACTIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT "IT WAS UNABLE TO PACE DURING USE." NO PATIENT COMPLICATIONS OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172238 SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION PACING CATHETER LDF EDWARDS LIFESCIENCES, PR PE075F5 59351241

Patients

Seq Age Sex Outcome Treatment
1