FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3071565 · Received April 22, 2013

Report

Report Number
2531779-2013-04812
Event Type
Malfunction
Date Received
April 22, 2013
Report Date
April 5, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 (B)(4) 2013 ¿ DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THERE WAS NO VISIBLE DAMAGE TO THE KEYPAD. THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND CONTAMINATION WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS. ADDITIONALLY, THE OK BUTTON CONTACT WAS FOUND TO BE INVERTED.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. IT WAS REPORTED THAT THE KEYPAD OK BUTTON IS INTERMITTENTLY UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172890 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 48 YR