FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3071563 · Received April 22, 2013

Report

Report Number
2531779-2013-04810
Event Type
Malfunction
Date Received
April 22, 2013
Report Date
April 4, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW UP #1 SUBMITTED: (B)(4) 2013 - DEVICE EVALUATION: ON EXAMINATION, THE KEYPAD WAS INTACT. ON TESTING UP ARROW, DOWN ARROW AND CONTRAST BUTTONS HAD INTERMITTENT RESPONSE. THE OK BUTTON WAS RESPONSIVE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER ALL THE KEYPAD BUTTONS. THE DISPLAY WAS NOTED TO BE FADED, DISCOLORED AND DIFFICULT TO READ. A TEST DISPLAY WAS ATTACHED TO THE PUMP AND ILLUMINATED PROPERLY WITHOUT DISCOLORATION.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. IT WAS REPORTED THE KEYPAD BUTTONS ARE INTERMITTENTLY UNRESPONSIVE. THE REPORTER STATED THIS CHANGE IN KEYPAD RESPONSE WAS IDENTIFIED AT LEAST 3 WEEKS AGO. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172316 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 19 YR