FDA Adverse Event Injury Summary report: N

ILS 29MM, CURVED

MDR report key: 3071443 · Received April 22, 2013

Report

Report Number
3005075853-2013-01885
Event Type
Injury
Date Received
April 22, 2013
Date of Event
March 25, 2013
Report Date
March 27, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K983536
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION. THE BREAKAWAY WASHER WAS PRESENT AND CUT AND THERE WERE NO STAPLES PRESENT, INDICATING THAT THE DEVICE ACHIEVED A FULL FIRING STROKE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE WAS USED FOR A PELVIC POUCH FORMATION. A SIGMOIDOSCOPY WAS PERFORMED AFTER USE OF THE DEVICE AND AN AIR LEK WAS NOTED AT THE STAPLE LINE, THEREFORE, A LOOP ILEOSTOMY HAD TO BE PERFORMED. ADDITIONAL INFORMATION HAS BEEN REQUESTED, NO RESPONSE HAS BEEN RECEIVED TO DATE, A SUPPLEMENTAL REPORT WILL BE SENT UPON RECEIPT OF ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
171792 ILS 29MM, CURVED STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK K4G242

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention