FDA Adverse Event Malfunction Summary report: N

EXCELSIOR SL-10 PRE-SHAPED 90 150CM

MDR report key: 3071404 · Received April 22, 2013

Report

Report Number
3008853977-2013-00108
Event Type
Malfunction
Date Received
April 22, 2013
Date of Event
April 4, 2013
Report Date
April 4, 2013
Manufacturer
STRYKER NEUROVASCULAR CORK
Product Code
DQY
PMA / PMN Number
K013789
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY RECORD REVIEW CONFIRMS THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. DURING VISUAL INSPECTION OF THE DEVICE IT WAS NOTED THERE WERE A SERIES OF KINKS AND FLATTENING AT 145 - 149 CM. A FUNCTIONAL TEST WAS PERFORMED AND RESISTANCE WAS FOUND WHEN INTRODUCING A MANDREL THROUGH THE CATHETER WHERE THE KINKS AND FLATTENING WERE NOTED. THE CATHETER WAS FLUSHED WITH WATER BUT DUE TO THE DAMAGE NO WATER WAS COMING OUT THE DISTAL END. NO HOLE WAS OBSERVED. A ROOT CAUSE OF HANDLING DAMAGE WILL BE ASSIGNED TO THIS COMPLAINT, AS IT WAS STATED THAT THE DEVICE WAS NOT USED THE LIKELY CAUSE OF THE DAMAGE NOTED COULD BE CAUSED DURING THE PREPARATION HANDLING OF THE DEVICE. THE MANUFACTURER HAS REVIEWED ALL INFORMATION AND DETERMINED THIS EVENT NO LONGER MEETS THE REQUIREMENT OF THE REPORTABLE EVENT FOR THE DEVICE IN QUESTION.

Description of Event or Problem · 1

DURING PREPRATION OF THE MICROCATHETER THE PHYSICIAN NOTED THERE WAS A HOLE BEHIND THE TIP OF THE CATHETER. THE PHYSICIAN HAD THE SAME ISSUE WITH A SECOND MICROCATHETER AND THE PROCEDURE WAS COMPLETED USING A THIRD MICROCATHETER. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

DURING PREPARATION OF THE MICROCATHETER THE PHYSICIAN NOTED THERE WAS A HOLE BEHIND THE TIP OF THE CATHETER. THE PHYSICIAN HAD THE SAME ISSUE WITH A SECOND MICROCATHETER AND THE PROCEDURE WAS COMPLETED USING A THIRD MICROCATHETER. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172462 EXCELSIOR SL-10 PRE-SHAPED 90 150CM CATHETER, PERCUTANEOUS DQY STRYKER NEUROVASCULAR CORK 15591373

Patients

Seq Age Sex Outcome Treatment
1 55 YR EXCELSIOR MICROCATHETER (BOSTON SCIENTIFIC)