FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3071352 · Received April 22, 2013

Report

Report Number
2531779-2013-04787
Event Type
Malfunction
Date Received
April 22, 2013
Report Date
March 27, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 07/02/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE PUMP¿S HISTORY SHOWED A TIME AND DATE RESET TO FACTORY DEFAULT FOLLOWING A REBOOT AT 11:56 PM ON (B)(6) 2013. THE TIME WAS THEN RESET TO 12:00 AM (B)(6) 2013. DURING TESTING, A TIME KEEPING ACCURACY TEST WAS PERFORMED. THE PUMP MAINTAINED TIME ACCURATELY DURING A 5 DAY DURATION TEST. THE PUMP WAS THEN POWERED OFF FOR AN HOUR AND WHEN THE PUMP WAS POWERED BACK ON, THE TIME AND DATE HAD RESET TO FACTORY DEFAULT. THE PUMP WAS OPENED AND THE INTERNAL BATTERY WAS FOUND TO BE LEAKING.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS (TIME/DATE ISSUE) ISSUE. THE REPORTER ALLEGED THE DATE WAS OFF BY ONE DAY, DISPLAYING (B)(6) INSTEAD OF (B)(6). THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172945 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 22 YR