ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-04780
- Event Type
- Malfunction
- Date Received
- April 22, 2013
- Report Date
- March 29, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE PUMP DISPLAY SCREEN WAS FOUND TO BE FADED AND DISCOLORED MAKING THE SCREEN DIFFICULT TO READ. THE DISPLAY SCREEN WAS REPLACED WITH A TEST SCREEN AND THE DISPLAY RETURNED TO NORMAL. UNRELATED TO THE DISPLAY ISSUE, THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED. THE PUMP BLACK BOX HISTORY REVEALED MULTIPLE UNEXPLAINED POWER EVENTS. DURING TESTING, A TEST BATTERY CAP WAS ABLE TO SECURE TO THE PUMP AND THE PUMP WAS ABLE TO EXERCISE FOR 24 HOURS WITHOUT POWER EVENTS OCCURRING. ALSO UNRELATED TO THE DISPLAY ISSUE, THE PUMP VIBRATORY ALARM WAS NOT RESPONDING; THE VIBRATION MOTOR WAS FOUND TO BE MISALIGNED ON THE PINS. ALSO UNRELATED TO THE DISPLAY ISSUE, EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PCB BOARD HAD FAILED. THE PUMP WOULD NOT RETAIN THE USER PROGRAMMED DATE AND TIME SETTINGS UPON REMOVAL OF THE PRIMARY AA BATTERY. WHEN A NEW AA BATTERY IS INSERTED THE PUMP DISPLAYS THE DEFAULT DATE AND TIME WHICH MUST BE MANUALLY CONFIRMED (OR RESET) BY THE USER IN ORDER TO PROCEED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).
THE PUMP WAS RETURNED TO ANIMAS. PRODUCT ANALYSIS EVALUATED THE PUMP AND FOUND A FADED AND DISCOLORED DISPLAY SCREEN. THIS REPORT IS MADE BASED ON THE RESULTS OF EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 172011 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |