Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECH FOUND THAT THE AUXILIARY POWER CORD HAD A LOOSE GROUND PRONG. HE REPLACED THE AUXILIARY POWER CORD AND THE BED FUNCTIONED TO SPECIFICATIONS.
Description of Event or Problem · 1
INFO RECEIVED INDICATED THE BED'S GROUND IMPEDANCE (RESISTANCE) IS OUT OF SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 144652 | TOTALCARE BED | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |