FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 3071299 · Received April 9, 2013

Report

Report Number
1824206-2013-02032
Event Type
Malfunction
Date Received
April 9, 2013
Date of Event
March 13, 2013
Report Date
March 13, 2013
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECH FOUND THAT THE AUXILIARY POWER CORD HAD A LOOSE GROUND PRONG. HE REPLACED THE AUXILIARY POWER CORD AND THE BED FUNCTIONED TO SPECIFICATIONS.

Description of Event or Problem · 1

INFO RECEIVED INDICATED THE BED'S GROUND IMPEDANCE (RESISTANCE) IS OUT OF SPECIFICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
144652 TOTALCARE BED AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1