FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC ACE

MDR report key: 3071215 · Received April 22, 2013

Report

Report Number
3005075853-2013-01910
Event Type
Malfunction
Date Received
April 22, 2013
Date of Event
April 2, 2013
Report Date
April 3, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED WITH THE DEVICE. THE REMAINING BLADE PORTION WAS SCRATCHED ¿ EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. THE DEVICE WAS ACTIVATED WITH THE GENERATOR AND AN ERROR CODE 5 WAS DISPLAYED. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN AN ERROR CODE 5 OR BLADE "LOCKOUT" LATER IN THE PROCEDURE, AND CONTINUED USAGE CAN RESULT IN A BROKEN BLADE

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE TITANIUM TIP BROKE. THE BROKEN PIECE WAS RETRIEVED. CHANGED TO ANOTHER ONE TO COMPETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT. ONE DEVICE WILL BE RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172520 ULTRACISION HARMONIC ACE INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. UNK UNK

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE