FDA Adverse Event Malfunction Summary report: N

AFFINITY 4 BIRTHING BED

MDR report key: 3071181 · Received April 9, 2013

Report

Report Number
1824206-2013-02048
Event Type
Malfunction
Date Received
April 9, 2013
Date of Event
March 13, 2013
Report Date
March 13, 2013
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE HEAD WALL UNIT HAD A FLOOR STEP INSTALLED WHICH CAUSED THE COMMUNICATION JUNCTION BOX AND SIDECOM CABLE TO BE DAMAGED WHEN BED WAS LOWERED WHILE IN THE TRENDELENBURG POSITION. THIS WAS A HILL-ROM HEAD WALL. THE TECHNICIAN STATED THE FLOOR STEP WAS REMOVED FROM THE WALL UNIT TO PREVENT FURTHER DAMAGE TO THE BED AND THE JUNCTION BOX AND SIDECOM CABLE WERE REPLACED TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THAT THE NURSE CALL DID NOT FUNCTION. NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
147933 AFFINITY 4 BIRTHING BED BIRTHING BED HDD HILL-ROM, INC. 3700

Patients

Seq Age Sex Outcome Treatment
1