FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 3071118
·
Received April 22, 2013
Report
- Report Number
- 0001811755-2013-00853
- Event Type
- Malfunction
- Date Received
- April 22, 2013
- Date of Event
- April 2, 2013
- Report Date
- April 2, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND TO BE BUILT UP THROUGHOUT THE INSIDE OF THE HANDPIECE. IT WAS ALSO FOUND THAT THERE WAS NO BOUNCE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CORE IMPACTION DRILL HEATED UP DURING A CASE. THERE WAS NO PATIENT OR USER INJURY REPORTED AND NO ADVERSE CONSEQUENCES ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 171723 | CORE IMPACTION DRILL | DRIVER, WIRE, AND BONE DRILL, | DZJ | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |