FDA Adverse Event Malfunction Summary report: N

LIGASURE ADVANCE PISTOL GRIP

MDR report key: 3070987 · Received April 1, 2013

Report

Report Number
1717344-2013-00241
Event Type
Malfunction
Date Received
April 1, 2013
Date of Event
March 12, 2013
Report Date
March 15, 2013
Manufacturer
COVIDIEN LP
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE KNIFE DID NOT ADVANCE WHEN THE TRIGGER WAS ACTIVATED AND IT WAS FOUND THAT THE KNIFE WAS HITTING THE BACK OF THE BLUE JAW SEAL PLATE. THE KNIFE EDGE WAS DAMAGED WHEN IT CONTACTED THE SEAL PLATE. THE KNIFE WAS STILL WITHIN THE JAWS, BUT THE WEBBING WAS BROKEN AND PROTRUDING FROM THE HINGE AREA. THE WEBBING MAY HAVE BROKEN IF FORCE WAS APPLIED TO THE TRIGGER AFTER THE KNIFE CONTACTED THE SEAL PLATE. A PRODUCTION SCREENING FIXTURE AND A DEEPER KNIFE SLOT IN THE JAW WERE PUT IN PLACE TO HELP MITIGATE THE OCCURRENCE OF THE KNIFE HITTING THE BACK OF THE SEAL PLATE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE KNIFE BLADE JAMMED DURING THE PROCEDURE. THE SURGEON MANUALLY OPENED THE DEVICE OFF THE PATIENT'S TISSUE AND OPENED ANOTHER TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY. THE DEVICE WAS RECEIVED AT COVIDIEN AND VISUAL INSPECTION DISCOVERED THE WEB PROTRUDING FROM THE HINGE OF THE JAW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
131295 LIGASURE ADVANCE PISTOL GRIP LIGASURE VESSEL SEALING SYSTEM GEI COVIDIEN LP 244998X

Patients

Seq Age Sex Outcome Treatment
1 UNK