FDA Adverse Event Malfunction Summary report: N

TURBOTRIMMER BLADE

MDR report key: 3070970 · Received April 22, 2013

Report

Report Number
1219602-2013-00091
Event Type
Malfunction
Date Received
April 22, 2013
Date of Event
March 26, 2013
Report Date
March 26, 2013
Manufacturer
SMITH & NEPHEW INC MANSFIELD MANUFACTURING SITE
Product Code
HAB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE RETURNED DEVICE WAS EVALUATED FOR SHEDDING AND IT WAS OBSERVED THAT THE OUTER BLADE RUNOUT WAS MEASURED TO BE .106". THIS IS 7 TIMES THE MAXIMUM DESIGN TOLERANCE AND RESULTED IN A MISALIGNMENT OF THE INNER AND OUTER BLADES. IT WAS ALSO OBSERVED THAT THERE WAS FRICTION PRESENT WHEN DISASSEMBLING THE INNER BLADE FROM THE OUTER BLADE. THE BLADE IS FOUND TO BE BENT, RESULTING IN SHEDDING. A BEND OF THIS MAGNITUDE IS TYPICALLY CAUSED BY EXCESSIVE LATERAL LOAD BEING APPLIED DURING USE. A REVIEW OF THE DEVICE HISTORY RECORDS AND QUALITY RECORDS ASSOCIATED WITH THIS MANUFACTURED LOT CONFIRMED THAT NO ADDITIONAL COMPLAINTS HAVE BEEN FILED AND THAT NO ABNORMALITIES WERE REPORTED WITH THIS PRODUCT DURING MANUFACTURE. (B)(6).

Description of Event or Problem · 1

WHILST SHAVING THE MENISCUS INSIDE THE KNEE JOINT, BITS OF METAL WERE COMING OFF THE SHAVER BLADE INTO THE JOINT, AND STICKING TO THE MENISCUS/AREA BEING RESECTED/SHAVED. SOME OF THE PIECES WERE ABLE TO BE REMOVED, BUT NOT ALL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172201 TURBOTRIMMER BLADE BOXED TURBOTRIMMER BL,4.5 MM/6 HAB SMITH & NEPHEW INC MANSFIELD MANUFACTURING SITE 50710451

Patients

Seq Age Sex Outcome Treatment
1 44 YR