FDA Adverse Event Malfunction Summary report: N

GYNECARE MORCELLEX* TISSUE MORCELLATOR

MDR report key: 3070895 · Received April 22, 2013

Report

Report Number
2210968-2013-04224
Event Type
Malfunction
Date Received
April 22, 2013
Date of Event
March 29, 2013
Report Date
April 4, 2013
Manufacturer
ETHICON, INC.
Product Code
HET
PMA / PMN Number
K100280
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. UPON EVALUATION, THE BLADE WAS NOT SEIZED AND THE DEVICE OPERATED AS INTENDED.

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA. THIS IS ONE OF TWO MEDWATCHES BEING SUBMITTED AS TWO DEVICES WERE INVOLVED IN THIS EVENT. SEE ALSO MEDWATCH 2210968-2013-04225. THE SAME PATIENT IS REPRESENTED IN EACH MEDWATCH.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A MYOMECTOMY ON (B)(6) 2013. DURING THE PROCEDURE, WHEN THE MYOMA WAS BEING TAKEN OUT, THE BLADE STOPPED ROTATING THOUGH THE SURGEON KEPT GRASPING THE TRIGGER. ANOTHER LIKE DEVICE WAS USED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172108 GYNECARE MORCELLEX* TISSUE MORCELLATOR LAPAROSCOPE, GYNECOLOGIC HET ETHICON, INC. NA MT217230

Patients

Seq Age Sex Outcome Treatment
1