FDA Adverse Event Injury Summary report: N

V40 FEM HEAD ORTHINOX 28+4

MDR report key: 3070779 · Received April 22, 2013

Report

Report Number
0002249697-2013-01345
Event Type
Injury
Date Received
April 22, 2013
Date of Event
March 23, 2013
Report Date
March 25, 2013
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
JDI
PMA / PMN Number
K011623
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CATALOG NUMBER AND LOT CODE WERE NOT PROVIDED. THE DEVICE WAS REPORTED AS AN UNKNOWN V40 ALUMINA HEAD. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF RECEIVED, WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

VISUAL INSPECTION REVEALED THE EXETER STEM IS BROKEN APPROXIMATELY AT THE BASE OF THE FEMORAL HEAD. THE FEMORAL HEAD IS STILL ATTACHED TO THE TRUNNION OF THE STEM. BASED ON THE PROVIDED INFORMATION, THE REPORTED FEMORAL HEAD DID NOT CONTRIBUTE TO THE EVENT.

Description of Event or Problem · 1

THE DISTRIBUTOR REPORTED ON BEHALF OF THE CUSTOMER THAT A REVISION FOR A BROKEN EXETER STEM TOOK PLACE ON (B)(6) 2013 IN (B)(6). THE PATIENT HAD PRIMARY DONE IN 2011 IN (B)(6). IT WAS REPORTED THAT THE PATIENT TWISTED HIS HIP WHILE CLOSING A GATE. THE STEM BROKE AT THE NECK AND THEREFORE, ALSO WORE AWAY PART OF THE POLY LINER. THE SURGEON HAD TO PERFORM A REVISION OF THE STEM, HEAD AND LINER.

Description of Event or Problem · 1

THE DISTRIBUTOR REPORTED ON BEHALF OF THE CUSTOMER THAT A REVISION FOR A BROKEN EXETER STEM TOOK PLACE ON (B)(6) 2013 IN (B)(6). THE PATIENT HAD PRIMARY DONE IN 2011 IN MERLIN PARK. IT WAS REPORTED THAT THE PATIENT TWISTED HIS HIP WHILE CLOSING A GATE. THE STEM BROKE AT THE NECK AND THEREFORE ALSO WORE AWAY PART OF THE POLY LINER. THE SURGEON HAD TO PERFORM A REVISION OF THE STEM, HEAD AND LINER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172942 V40 FEM HEAD ORTHINOX 28+4 IMPLANT JDI STRYKER ORTHOPAEDICS-MAHWAH G2835367

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| O| R