Report
- Report Number
- 1415939-2013-00178
- Event Type
- Malfunction
- Date Received
- April 22, 2013
- Report Date
- March 31, 2013
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- NIG
- PMA / PMN Number
- K052000
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
EVALUATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, A SEARCH FOR SIMILAR COMPLAINTS, RELEASE TESTING DATA REVIEW, SPECIFICITY TESTING, AND A REVIEW OF LABELING. A REVIEW OF CUSTOMER COMPLAINTS RECEIVED TO-DATE DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY RELATED TO DISCREPANT PATIENT RESULTS WITH ARCHITECT CA19-9XR, LIST 2K91-25, LOT 18067M500. ACCURACY TESTING WAS COMPLETED USING AN INTERNAL CA19-9XR PANEL AND RETAINED KITS OF REAGENT LOT 18067M500, WHICH MET ACCEPTANCE CRITERIA. THE INVESTIGATION RESULTS SHOW THAT THERE IS NO PRODUCT DEFICIENCY AND THAT THE ARCHITECT CA 19-9XR REAGENTS ARE PERFORMING AS INTENDED. FURTHER INVESTIGATION IS NOT REQUIRED. BASED ON THE RESULTS OF THIS EVALUATION, THE ARCHITECT CA19-9XR ASSAY, LIST 2K91-25, LOT 18067M500, IS PERFORMING AS INTENDED AND NO PRODUCT DEFICIENCY OR MALFUNCTION WAS IDENTIFIED.
THE CUSTOMER STATES THAT THE ARCHITECT ANALYZER IS GENERATING FALSELY ELEVATED CA19-9 RESULTS ON ONE PATIENT SAMPLE. TESTING WAS RUN AND GENERATED THE FOLLOWING RESULTS: 33NG/ML; 5 HOURS LATER -143 NG/ML; AND THEN 43 AND 34 NG/ML. THE SAMPLE WAS SENT TO A REFERENCE LAB WHERE THE FOLLOWING RESULTS WERE GENERATED: 47.18, 40.74, AND 40.47 NG/ML. THERE WAS NO ADDITIONAL PATIENT INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 172007 | ARCHITECT CA 19-9XR | NIG | ABBOTT LABORATORIES | 18067M500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCHITECT I2000SR ANALYZER, LIST # 03M74-01| SERIAL # (B)(4)| ARCHITECT I2000SR ANALYZER, LIST # 03M74-01| SERIAL # (B)(4) |