FDA Adverse Event Malfunction Summary report: N

ACCURUS 800CS

MDR report key: 3070643 · Received April 17, 2013

Report

Report Number
2028159-2013-00687
Event Type
Malfunction
Date Received
April 17, 2013
Date of Event
March 15, 2013
Report Date
March 20, 2013
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K911808
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY SERVICE REPRESENTATIVE EXAMINED THE SYSTEM AND WAS UNABLE TO CONFIRM THE REPORTED EVENT. THERE WERE NO RELATED SYSTEM MESSAGES IN THE EVENT LOG. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. THERE WAS NO SAMPLE RETURNED FOR EVALUATION AND NO ADDITIONAL INFORMATION PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID NOT INDICATE AND ADDITIONAL SIMILAR REPORT FOR THIS SYSTEM. THE ROOT CAUSE CANNOT BE DETERMINED CONCLUSIVELY. (B)(4).

Description of Event or Problem · 1

A NURSE REPORTED THAT A SURGEON EXPERIENCED POOR PRESSURE AND THE PATIENT HAD A SOFT EYE DURING A VITREORETINAL PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED, HOWEVER, NONE HAS BEEN PROVIDED TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
166290 ACCURUS 800CS PHACOFRAGMENTATION SYSTEM HQC ALCON - IRVINE TECHNOLOGY CENTER ACCURUS NA