EX LIFT
Report
- Report Number
- 2183887-2013-00001
- Event Type
- Injury
- Date Received
- April 12, 2013
- Date of Event
- December 17, 2012
- Report Date
- April 12, 2013
- Manufacturer
- EZ WAY, INC.
- Product Code
- FNG
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
INITIAL REPORTER DID NOT NOTIFY EZ WAY OF THE INCIDENT. EZ WAY RECEIVED COPY OF THE MDR SUBMITTED BY THE FACILITY FROM THE FDA. IN RESPONSE TO A NOTIFICATION ON (B)(4) 2005 BY THE OEM ACTUATOR MFR, SKF, THAT THESE ACTUATORS SHOULD BE REPLACED AFTER 4 YEARS, EZ WAY ISSUED THREE CONSECUTIVE LETTERS SENT (B)(4) 2005, (B)(4)2006 TO ALL CUSTOMER OWNING EZ LIFTS PREVIOUSLY MFG WITH THIS BRAND OF ACTUATOR STATING THE NEED TO REPLACE SKF ACTUATORS AFTER 4 YEARS, ALONG WITH A MAINTENANCE CHECKLIST SPECIFYING ACTUATOR MAINTENANCE GUIDELINES TO TAKE PLACE ON A MONTHLY BASIS. EZ WAY HAS NOT USED THIS ACTUATOR FOR PRODUCTION OR AS A REPLACEMENT PART SINCE (B)(4) 1999.
DURING A TRANSFER, THE FLANGE ATTACHMENT FOR THE ACTUATOR'S OUTER TUBE CAME UNATTACHED FROM THE GAR BOX AT THE ATTACHMENT POINT CAUSING THE LEAD SCREW TO FALL. THE ACTUATOR INVOLVED WAS THE ORIGINAL ACTUATOR THAT WAS SOLD WITH THE MACHINE ON (B)(6) 1997. INAPPROPRIATE HANDLING OF THE ACTUATOR, ABNORMAL USE, AND LACK OF PROPER SERVICING OF THE ACTUATOR PRESENTS A RISK OF DAMAGE AND MATERIAL FATIGUE. THE ACTUATOR MFR, SKF/MAGNETIC CORPORATION, RECOMMENDS THAT ACTUATORS THAT HAVE BEEN IN SERVICE FOR MORE THAN FOUR YEARS BE REPLACED. THE ACTUATOR INVOLVED IN THE INCIDENT WAS THE ORIGINAL ACTUATOR SOLD WITH THE EZ LIFT IN 1997 AND HAD BEEN IN USE FOR 16 YEARS. EZ WAY PREVIOUSLY CONTACTED THE USER FACILITY THREE TIMES VIA LETTER AND SIX TIMES VIA PHONE STATING THE NEED TO REPLACE THEIR SKF ACTUATOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 157933 | EX LIFT | PATIENT LIFT | FNG | EZ WAY, INC. | 796 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Other |