FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3070534 · Received April 17, 2013

Report

Report Number
8020893-2013-00875
Event Type
Malfunction
Date Received
April 17, 2013
Date of Event
March 19, 2013
Report Date
March 20, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TH
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD AN ERRATIC SCREEN. THERE WAS NO PATIENT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) CPU PCB. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
165195 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN 840

Patients

Seq Age Sex Outcome Treatment
1