FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3070524 · Received April 17, 2013

Report

Report Number
8020893-2013-00880
Event Type
Malfunction
Date Received
April 17, 2013
Date of Event
March 18, 2013
Report Date
March 18, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR IS INOPERABLE. IT IS UNK IF THERE WAS PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE BREATH DELIVERY UNIT (BDU) PCB. THE UNIT PASSED EXTENDED SELF TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
165614 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN 840

Patients

Seq Age Sex Outcome Treatment