FDA Adverse Event Injury Summary report: N

DEXTRUS 4137

MDR report key: 3070484 · Received April 22, 2013

Report

Report Number
1028232-2013-01107
Event Type
Injury
Date Received
April 22, 2013
Date of Event
February 21, 2013
Report Date
April 9, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT, THE LEAD WAS SUBJECTED TO AN EXTENSIVE ANALYSIS. THE PERFORMANCE OF THE DEVICE UNDER COMPLAINT WAS SCRUTINIZED, INCLUDING A VISUAL, ELECTRICAL AND MECHANICAL INSPECTION. THE VISUAL INSPECTION SHOWED CUTTINGS IN THE INSULATION AND DEFORMATIONS OF THE CONDUCTOR COIL WHICH OCCURRED MOST LIKELY DURING SURGERY. DURING FURTHER ANALYSIS, NO DEVIATIONS WERE NOTED WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION AS MENTIONED IN THE COMPLAINT DESCRIPTION. THE LEAD PROVED TO BE WITHIN SPECIFICATIONS AND FLAWLESS THROUGHOUT ITS INSPECTION. IN SUMMARY, THERE WAS NO SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE. BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ASSOCIATED WITH THE MANUFACTURE OF THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE RIGHT VENTRICULAR LEAD DISLODGED AND THE PATIENT HAD SEVERE TRICUSPID REGURGITATION. THE PHYSICIAN PERFORMED SURGICAL INTERVENTION TO REMOVE THE RV LEAD AND REPLACED IT WITH A HEAVIER LEAD. THIS LEAD IS NOW OUT OF SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
173192 DEXTRUS 4137 PACER LEAD NVN BIOTRONIK SE & CO. KG 358764

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization