FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK SPIRIT COMBO

MDR report key: 3070444 · Received April 22, 2013

Report

Report Number
2183996-2013-00718
Event Type
Malfunction
Date Received
April 22, 2013
Date of Event
April 1, 2013
Report Date
July 29, 2013
Manufacturer
ROCHE DIABETES CARE AG
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION THE COMPLAINT CANNOT BE VERIFIED DUE TO MISHANDLING OF THE PRODUCT. RESULT THE INSULIN PUMP IS DAMAGED DUE TO A HARD MECHANICAL IMPACT. DUE TO SEVERE DAMAGES ON THE MECHANICAL AND ELECTRONIC COMPONENTS THE INSULIN PUMP TRIGGERED ERROR MESSAGES. E6 WERE FOUND IN THE PUMP HISTORY. THE INSULIN PUMP SHOULDN'T BE EXPOSED TO HIGH MECHANICAL FORCES. THE VIBRA DOES NOT FUNCTION. THE VIBRA SLIPPED OUT OF HIS HOLDER AND TOUCHED THE FORCE SENSOR (HOLDER WALL). DUE TO THIS PROBLEM THE VIBRA GET BLOCKED AND DID NOT FUNCTION AS INTENDED. AN EXTERNAL MECHANICAL INFLUENCE (HARD IMPACT / DROP) CAUSED THE DAMAGE. THE PROBLEM MENTIONED BY THE CUSTOMER IS RELATED TO A HANDLING ISSUE.

Additional Manufacturer Narrative · 1

THIS INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

PATIENT REPORTED RECEIVING THE MESSAGE "MECHANICAL ERROR" ON THE INFUSION DEVICE. PATIENT STATED HE CHANGED THE BATTERY AND INFUSION SET, BUT THE MESSAGE CAME UP AGAIN. PATIENT REPORTED THE VIBRATION MECHANISM HAS STOPPED WORKING. NO FURTHER INFORMATION IS AVAILABLE. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. REQUESTED RETURN OF THE ALLEGED INFUSION DEVICE FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
173011 ACCU-CHEK SPIRIT COMBO INSULIN INFUSION PUMP LZG ROCHE DIABETES CARE AG 00700006863 NA

Patients

Seq Age Sex Outcome Treatment
1