LEAD MODEL 302
Report
- Report Number
- 1644487-2013-01100
- Event Type
- Malfunction
- Date Received
- April 22, 2013
- Date of Event
- December 22, 2012
- Report Date
- March 25, 2013
- Manufacturer
- CYBERONICS INC
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NURSE
Narratives
DATE OF EVENT, CORRECTED DATA: THE CORRECT EVENT DATE FOR HIGH LEAD IMPEDANCE IS PROVIDED PER REVIEW OF INTERNAL GENERATOR DATA. DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
DEVICE FAILURE IS SUSPECTED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
REPORTER INDICATED THAT A LEAD DISCONTINUITY WAS NOTED DURING VNS GENERATOR REPLACEMENT SURGERY ON (B)(6) 2013 FOR END OF SERVICE, AND A NEW LEAD AND GENERATOR WERE IMPLANTED. ADDITIONAL FOLLOW UP WITH THE REPORTER REVEALED THE LEAD DISCONTINUITY WAS FIRST NOTED ON (B)(6) 2013 WITH HIGH LEAD IMPEDANCE >10,000 OHMS WITH DIAGNOSTICS TESTING, BUT X-RAYS DID NOT VISUALIZE A LEAD FRACTURE. NO TRAUMA OR DEVICE MANIPULATION OCCURRED. THE X-RAYS WILL NOT BE RELEASED TO THE MANUFACTURER. THE EXPLANTED VNS LEAD AND GENERATOR WERE RETURNED ON (B)(4) 2013 AND ARE PENDING ANALYSIS.
PRODUCT ANALYSIS WAS COMPLETED ON THE RETURNED VNS GENERATOR AND LEAD. ANALYSIS OF THE GENERATOR DID NOT IDENTIFY ANY ANOMALIES THAT COULD AFFECT DEVICE FUNCTION. THE GENERATOR WAS AT A NEAR END OF SERVICE CONDITION (NEOS). REVIEW OF THE GENERATOR¿S INTERNAL MEMORY IDENTIFIED THAT AN IMPEDANCE CHANGE FROM 2353 OHMS TO 13059 OHMS OCCURRED ON (B)(6) 2012. ANALYSIS OF THE LEAD CONFIRMED A BREAK WAS IDENTIFIED IN THE POSITIVE COIL. SCANNING ELECTRON MICROSCOPY IMAGES OF THE POSITIVE COIL SHOW THAT PITTING OR ELECTRO-ETCHING CONDITIONS HAVE OCCURRED AT THE BREAK LOCATION. HOWEVER, DUE TO MECHANICAL DISTORTION (SMOOTHED SURFACES), PITTING AND/OR SURFACE CONTAMINATION THE FRACTURE MECHANISM CANNOT BE DETERMINED. ALSO, A SINGLE BROKEN STRAND WAS NOTED IN THE NEGATIVE COIL. THE SILICONE TUBING OF THE POSITIVE COIL IS ABRADED OPEN. DRY BODY FLUIDS WERE NOTED INSIDE THE INNER SILICONE TUBING. NOTE THAT SINCE THE ELECTRODE ARRAY PORTION WAS NOT RETURNED FOR ANALYSIS, AN EVALUATION AND RESULTING COMMENTARY CANNOT BE MADE ON THAT PORTION OF THE LEAD. OTHER THAN THE ABOVE MENTIONED OBSERVATIONS AND TYPICAL WEAR AND EXPLANT RELATED OBSERVATIONS, NO OTHER ANOMALIES WERE IDENTIFIED IN THE RETURNED LEAD PORTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 171153 | LEAD MODEL 302 | LEAD | LYJ | CYBERONICS INC | 302-20 | 200936 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 YR |