EASYTRAK 2
Report
- Report Number
- 2124215-2013-06496
- Event Type
- Injury
- Date Received
- April 22, 2013
- Date of Event
- August 1, 2012
- Report Date
- March 12, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA
- Reporter Occupation
- PHYSICIAN
Narratives
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT APPROXIMATELY FOUR MONTHS AFTER THE PATIENT'S LAST GENERATOR CHANGE-OUT, THE LEFT VENTRICULAR (LV) PACING IMPEDANCE WAS GREATER THAN 2000 OHMS. ALL PACING VECTORS WERE TESTED, THE LEAD WAS NON-FUNCTIONAL, AND THE PHYSICIAN SUSPECTED THE LEAD WAS FRACTURED. RECENTLY, IT WAS DECIDED TO REPLACE THE LEAD AND THE PATIENT UNDERWENT SURGICAL INTERVENTION. THE LEAD WAS EVALUATED WITH THE GENERATOR AND PACING SYSTEM ANALYZER (PSA) AND FOUND TO HAVE NO CAPTURE IN THE BIPOLAR, TIP TO CAN, AND TIP TO RV PACING CONFIGURATIONS. FLUOROSCOPY EVALUATION WAS PERFORMED AND THE PHYSICIAN WAS UNABLE TO IDENTIFY A FRACTURE. THE LEAD WAS SURGICALLY ABANDONED AND REPLACED, AND THE GENERATOR REMAINS IN SERVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 172899 | EASYTRAK 2 | IMPLANTABLE LEAD | OJX | CPI - DEL CARIBE | 4543 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Hospitalization| L| R | 5076| H217| N119| 4543| 6947| 1853 |