FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 3070319 · Received April 22, 2013

Report

Report Number
2024168-2013-02468
Event Type
Injury
Date Received
April 22, 2013
Date of Event
February 27, 2013
Report Date
March 28, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. ESTIMATED DATE OF OCCURRENCE, EXACT DATE WAS NOT KNOWN. THE OTHER PROGLIDE DEVICE REFERENCED, IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED EVENT WAS NOT CONFIRMED. ANALYSIS OF THE RETURNED DEVICE COULD NOT VERIFY OR CONFIRM THE REPORTED CUFF MISS BECAUSE ONLY THE DEVICE BODY WAS RETURNED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. NINE UNUSED STERILE PROGLIDE DEVICES WITH THE LOT NUMBER, 30215K1, WERE RETURNED FOR EVALUATION. FUNCTIONAL TESTING WAS PERFORMED AND THE DEVICES PASSED WITH ACCEPTABLE RESULTS. NO MANUFACTURING OR ABNORMAL OBSERVATIONS WERE DETECTED. A CAUSE FOR THE COMPLAINT DEVICE REPORTED EXPERIENCE COULD NOT BE DETERMINED BASED ON THE INVESTIGATION FINDINGS FROM THE TESTED SAMPLES. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE WITH A 6F SHEATH, AFTER AN UNSPECIFIED PROCEDURE. REPORTEDLY, NO SUTURE WAS PRESENT WHEN THE PLUNGER WAS PULLED, A CUFF MISS OCCURRED. A SECOND PROGLIDE WAS USED WITH THE SAME RESULTS. A NON-ABBOTT DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE PHYSICIAN IS REPORTEDLY TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172697 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 30124J1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 6F