PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2013-02468
- Event Type
- Injury
- Date Received
- April 22, 2013
- Date of Event
- February 27, 2013
- Report Date
- March 28, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. ESTIMATED DATE OF OCCURRENCE, EXACT DATE WAS NOT KNOWN. THE OTHER PROGLIDE DEVICE REFERENCED, IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED EVENT WAS NOT CONFIRMED. ANALYSIS OF THE RETURNED DEVICE COULD NOT VERIFY OR CONFIRM THE REPORTED CUFF MISS BECAUSE ONLY THE DEVICE BODY WAS RETURNED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. NINE UNUSED STERILE PROGLIDE DEVICES WITH THE LOT NUMBER, 30215K1, WERE RETURNED FOR EVALUATION. FUNCTIONAL TESTING WAS PERFORMED AND THE DEVICES PASSED WITH ACCEPTABLE RESULTS. NO MANUFACTURING OR ABNORMAL OBSERVATIONS WERE DETECTED. A CAUSE FOR THE COMPLAINT DEVICE REPORTED EXPERIENCE COULD NOT BE DETERMINED BASED ON THE INVESTIGATION FINDINGS FROM THE TESTED SAMPLES. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE WITH A 6F SHEATH, AFTER AN UNSPECIFIED PROCEDURE. REPORTEDLY, NO SUTURE WAS PRESENT WHEN THE PLUNGER WAS PULLED, A CUFF MISS OCCURRED. A SECOND PROGLIDE WAS USED WITH THE SAME RESULTS. A NON-ABBOTT DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE PHYSICIAN IS REPORTEDLY TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 172697 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 30124J1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | SHEATH: 6F |