FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 3070269
·
Received April 16, 2013
Report
- Report Number
- 2518422-2013-00590
- Event Type
- Malfunction
- Date Received
- April 16, 2013
- Date of Event
- March 29, 2013
- Report Date
- March 29, 2013
- Manufacturer
- RESPIRONICS INC.
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE VENTILATOR WAS RETURNED TO A THIRD PARTY SERVICE CENTER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S POWER SUPPLY WAS REPLACED TO ADDRESS THE ISSUE. THE MALFUNCTION OF THE POWER SUPPLY WOULD RESULT IN THE FAILURE OF THE DEVICE TO OPERATE ON AC POWER OR TO CHARGE ITS INTERNAL BATTERY. THIS PARTICULAR MALFUNCTION FOR THE POWER SUPPLY HAS ONLY BEEN REPORTED ON DEVICES THAT ARE BEING SOLD INTERNATIONALLY FOR USE WITH 220 VOLT POWER SOURCES.
Description of Event or Problem · 1
THE MANUFACTURER RECEIVED INFORMATION FROM A THIRD PARTY SERVICE CENTER ALLEGING A VENTILATOR WOULD NOT POWER ON. THERE WAS NO HARM OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 164085 | TRILOGY 100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS INC. | 1054096 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |