FDA Adverse Event Malfunction Summary report: N

PERMOBIL C350 CORPUS

MDR report key: 3070241 · Received April 1, 2013

Report

Report Number
1221084-2013-00007
Event Type
Malfunction
Date Received
April 1, 2013
Date of Event
March 4, 2013
Report Date
March 29, 2013
Manufacturer
PERMOBIL AB
Product Code
ITI
PMA / PMN Number
K071650
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE WELDS ON THE BOTTOM PLATE OF THE SEAT TUBE HAVE BROKEN, THIS ALLOWED THE SEAT TO MOVE EXCESSIVLEY WHICH CONTIBUTED TO THIS INSTANCE. THE WELDING PARAMETERS HAVE BEEN UPDATED FOR THIS PART TO PROVIDE BETTER LONGEVITY AND QUALITY. THE NEWER PART HAS BEEN INSTALLED ON THIS CHAIR.

Description of Event or Problem · 1

MANUFACTURER RECEIVED A REPORT THAT THE CLIENT'S POWER WHEELCHAIR SEAT SEPARATED FROM THE CHASSIS WHEN HE WAS GOING DOWN A RAMP. CLIENT DID NOT SEEK MEDICAL ATTENTION. MANUFACTURER HAS NOT RECEIVED FURTHER INFORMATION OR INSPECTED THE WHEELCHAIR AS OF THE DATE OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
133188 PERMOBIL C350 CORPUS POWER WHEELCHAIR ITI PERMOBIL AB

Patients

Seq Age Sex Outcome Treatment
1