FDA Adverse Event
Malfunction
Summary report: N
PERMOBIL C350 CORPUS
MDR report key: 3070241
·
Received April 1, 2013
Report
- Report Number
- 1221084-2013-00007
- Event Type
- Malfunction
- Date Received
- April 1, 2013
- Date of Event
- March 4, 2013
- Report Date
- March 29, 2013
- Manufacturer
- PERMOBIL AB
- Product Code
- ITI
- PMA / PMN Number
- K071650
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE WELDS ON THE BOTTOM PLATE OF THE SEAT TUBE HAVE BROKEN, THIS ALLOWED THE SEAT TO MOVE EXCESSIVLEY WHICH CONTIBUTED TO THIS INSTANCE. THE WELDING PARAMETERS HAVE BEEN UPDATED FOR THIS PART TO PROVIDE BETTER LONGEVITY AND QUALITY. THE NEWER PART HAS BEEN INSTALLED ON THIS CHAIR.
Description of Event or Problem · 1
MANUFACTURER RECEIVED A REPORT THAT THE CLIENT'S POWER WHEELCHAIR SEAT SEPARATED FROM THE CHASSIS WHEN HE WAS GOING DOWN A RAMP. CLIENT DID NOT SEEK MEDICAL ATTENTION. MANUFACTURER HAS NOT RECEIVED FURTHER INFORMATION OR INSPECTED THE WHEELCHAIR AS OF THE DATE OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 133188 | PERMOBIL C350 CORPUS | POWER WHEELCHAIR | ITI | PERMOBIL AB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |