FDA Adverse Event
Malfunction
Summary report: N
TRILOGY O2
MDR report key: 3070231
·
Received April 16, 2013
Report
- Report Number
- 2518422-2013-00593
- Event Type
- Malfunction
- Date Received
- April 16, 2013
- Date of Event
- March 29, 2013
- Report Date
- March 29, 2013
- Manufacturer
- RESPIRONICS INC., SLEEP & HOME RESPIRATORY GROUP
- Product Code
- CBK
- PMA / PMN Number
- K093905
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MFR FOR EVAL AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S OXYGEN BLENDING MODULE BOARD WAS REPLACED TO ADDRESS THE ISSUE. THE DEVICE FAILED A STEP DURING TESTING. THE DEVICE'S SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUE.
Description of Event or Problem · 1
THE MFR RECEIVED INFO ALLEGING A "SERVICE REQUIRED" ALARM CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 164016 | TRILOGY O2 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS INC., SLEEP & HOME RESPIRATORY GROUP | 1040004 |