FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 3070213 · Received April 22, 2013

Report

Report Number
1644487-2013-01097
Event Type
Malfunction
Date Received
April 22, 2013
Date of Event
September 13, 2008
Report Date
March 24, 2013
Manufacturer
CYBERONICS
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF PROGRAMMING HISTORY.

Description of Event or Problem · 1

DURING REVIEW OF PROGRAMMING HISTORY IT WAS OBSERVED THAT A INCOMPLETE SYSTEMS DIAGNOSTICS OCCURRED THAT LEFT THE PATIENT PROGRAMMING TO UNINTENTIONAL SETTINGS. THERE WAS NO FINAL INTERROGATION PRIOR TO THE PATIENT LEAVING THE OFFICE AND THE SETTING CHANGED WAS NOT NOTICED AND CORRECTED UNTIL THE NEXT APPOINTMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172982 PROGRAMMING SOFTWARE PROGRAMMING SOFTWARE LYJ CYBERONICS MODEL 250 523500

Patients

Seq Age Sex Outcome Treatment
1 58 YR