FDA Adverse Event Malfunction Summary report: N

STERNAL ZIPFIX WITH NEEDLE STERILE

MDR report key: 3070106 · Received April 22, 2013

Report

Report Number
8030965-2013-10978
Event Type
Malfunction
Date Received
April 22, 2013
Date of Event
December 15, 2011
Report Date
December 15, 2011
Manufacturer
SYNTHES GMBH
Product Code
JDQ
PMA / PMN Number
K110789
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. (B)(4). PLACEHOLDER.

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. INVESTIGATION COULD NOT BE COMPLETED AND NO CONCLUSION COULD BE DRAWN AS NO DEVICE WAS RETURNED AND NO LOT NUMBER WAS PROVIDED. PLACEHOLDER.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STERNAL CLOSURE FOLLOWING AN OSTEOTOMY PROCEDURE, THE ZIPFIX BROKE WHILE THE SURGEON WAS TRYING TO WORK THE NEEDLE THROUGH THE SECOND HALF OF THE STERNUM. DUE TO THE SHORTNESS OF THE NEEDLE AND THICKNESS OF THE INTERCOSTAL SPACE, THE SURGEON WAS REALLY TRYING TO PUSH THE NEEDLE THROUGH AND IT BROKE OFF WHERE THE LITTLE NOTCH IS BELOW THE NEEDLE. SURGEON IMPLANTED FIVE ZIPFIXES, AND OUT OF THE FIVE TWO BROKE. DEVICES WERE STILL ABLE TO BE IMPLANTED TO THE PATIENT. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
171894 STERNAL ZIPFIX WITH NEEDLE STERILE JDQ SYNTHES GMBH

Patients

Seq Age Sex Outcome Treatment
1