FDA Adverse Event Malfunction Summary report: N

MAESTRO LARGE FIXED DURAGUARD

MDR report key: 3070098 · Received April 22, 2013

Report

Report Number
0001811755-2013-00850
Event Type
Malfunction
Date Received
April 22, 2013
Date of Event
April 4, 2013
Report Date
April 4, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING FAILURE ANALYSIS, THE REPORTED EVENT OF OVERHEATING WAS DUPLICATED AND CONFIRMED. VISUAL INSPECTION DURING DISASSEMBLY CONFIRMED DEBRIS AFFECTING THE BEARINGS AND FRONT BEARING SLEEVE. DEBRIS CAN CAUSE THE BEARINGS TO WEAR AND GRADUALLY LOSE THEIR ROTATIONAL PROPERTIES THROUGH DEGRADATION, WHICH CAN CAUSE FRICTIONAL HEAT. THE DEVICE WAS DISCARDED BY THE MANUFACTURER FOLLOWING EVALUATION. DURING EVALUATION, IT WAS CONFIRMED THAT THE DURAGUARD WAS NOT BENT. THEREFORE, CORRECTIONS HAVE BEEN MADE TO EXECUTIVE SUMMARY AND DEVICE CODES TO REFLECT THIS FINDING.

Additional Manufacturer Narrative · 1

THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; IT IS NOT POSSIBLE TO DETERMINE THE CAUSE OF THE REPORTED EVENT WITHOUT AN EVALUATION OF THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY, THE DURA GUARD WAS OVERHEATING. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY, THE DURA GUARD WAS OVERHEATING AND WAS BENT. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.