MAESTRO LARGE FIXED DURAGUARD
Report
- Report Number
- 0001811755-2013-00850
- Event Type
- Malfunction
- Date Received
- April 22, 2013
- Date of Event
- April 4, 2013
- Report Date
- April 4, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HBB
- PMA / PMN Number
- K041754
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
DURING FAILURE ANALYSIS, THE REPORTED EVENT OF OVERHEATING WAS DUPLICATED AND CONFIRMED. VISUAL INSPECTION DURING DISASSEMBLY CONFIRMED DEBRIS AFFECTING THE BEARINGS AND FRONT BEARING SLEEVE. DEBRIS CAN CAUSE THE BEARINGS TO WEAR AND GRADUALLY LOSE THEIR ROTATIONAL PROPERTIES THROUGH DEGRADATION, WHICH CAN CAUSE FRICTIONAL HEAT. THE DEVICE WAS DISCARDED BY THE MANUFACTURER FOLLOWING EVALUATION. DURING EVALUATION, IT WAS CONFIRMED THAT THE DURAGUARD WAS NOT BENT. THEREFORE, CORRECTIONS HAVE BEEN MADE TO EXECUTIVE SUMMARY AND DEVICE CODES TO REFLECT THIS FINDING.
THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; IT IS NOT POSSIBLE TO DETERMINE THE CAUSE OF THE REPORTED EVENT WITHOUT AN EVALUATION OF THE DEVICE.
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY, THE DURA GUARD WAS OVERHEATING. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY, THE DURA GUARD WAS OVERHEATING AND WAS BENT. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 171640 | MAESTRO LARGE FIXED DURAGUARD | MOTOR, DRILL, PNEUMATIC | HBB | STRYKER INSTRUMENTS-KALAMAZOO | 06335 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |