FDA Adverse Event Malfunction Summary report: N

DRILL 1882900 MINI TREPHINATION 2MM

MDR report key: 3066565 · Received April 18, 2013

Report

Report Number
1045254-2013-00338
Event Type
Malfunction
Date Received
April 18, 2013
Date of Event
March 21, 2013
Report Date
March 22, 2013
Manufacturer
MEDTRONIC XOMED, INC.
Product Code
EQJ
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS PRODUCT IS USED FOR THERAPEUTIC PURPOSES. (B)(4). THE DEVICE WAS RETURNED AND PRODUCT EVALUATION IS CURRENTLY UNDERWAY. THE PRELIMINARY ANALYSIS CONFIRMED THE DRILL BIT TIPS WERE DETACHED. RESULTS: RESULTS PENDING COMPLETION OF EVALUATION. (B)(4).

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: UPON EVALUATION, THE SHAFTS ON BOTH OF THE RETURNED SAMPLES WERE BROKEN CLOSE TO THE DISTAL END. THE BROKEN PIECES MEASURED APPROXIMATELY 1.7CMS. ONE SAMPLE SHOWED CIRCULAR WEAR MARKS, POSSIBLY INDICATING EXCESSIVE USE/HANDLING OF THE DEVICE BY THE CUSTOMER AND LONG RUN TIME. THE SECOND SAMPLE DID NOT EXHIBIT ANY OBVIOUS MARKINGS ON THE SHAFT. THE BREAKAGE APPEARED TO HAVE OCCURRED AT THE SHARP CORNER / BREAK EDGE. THE OUTER DIAMETER OF THE SHAFT MEASURED APPROXIMATELY 0.0920" WHICH MEETS THE REQUIRED SPECIFICATION OF 0.092"+.0006-.0000. UNDER MAGNIFICATION, THE DRILL BIT TIPS DID NOT REVEAL ANY OBVIOUS DAMAGE OR FRAGMENTATION. THE CUSTOMER¿S COMPLAINT WAS CONFIRMED; THE EXACT ROOT CAUSE OF THE COMPLAINT WAS NOT CONFIRMED, THE MOST LIKELY / PROBABLE CAUSE OF THE ISSUE APPEARS TO BE RELATED WITH CUSTOMER MISUSE / MISHANDLING OF THE PRODUCT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE TIPS OF TWO DRILL BITS BROKE OFF ONCE THE HANDPIECE WAS ACTIVATED. THERE WAS NO PATIENT INJURY.